Thông số máy Fluke 54-2
- Dual Input Digital Thermometer including:
- Impact absorbing holster
- Two 80PK-1 bead probe thermocouples
- Temperature accuracy
Above -100 °C:
J, K, T, E, and N-type: ±[0.05% + 0.3°C]*
R and S-type: ±[0.05% + 0.4°C]*
Below -100 °C:
J, K, E, and N-types: ±[0.20% + 0.3°C]*
T-type: ±[0.50% + 0.3°C]
J-type: -210 °C to 1200 °C
K-type: -200 °C to 1372 °C
T-type: -250 °C to 400 °C
E-type: -150 °C to 1000 °C
N-type: -200 °C to 1300 °C *
R and S-type: 0 °C to 1767 °C *
- Temperature scale: ITS-90
- Applicable standards: NIST-175
- Display Resolution:
0.1 °C, 0.1 K < 1000
1°C, 1 K >=1000
- Note:* Only the Fluke Models 53 II and 54 II thermometers are capable of measuring N, R, and S-type thermocouples.
- Environmental Specifications
- Operating Temperature: -10 °C to 50 °C
- Storage Temperature: -40 °C to 60 °C
- Humidity (Without Condensation):
0% to 90%; 0 °C to 35 °C
0% to 70%; 0 °C to 50 °C
- Safety Specifications
- Overvoltage category: CSA C22.2 No. 1010.1 1992; EN 61010 Amendments 1,2
- Agency Approvals : CE, CSA, TÜV (pending)
- Mechanical & General Specifications
- Size: 173 mm L x 86 mm W x 38 mm
- Weight: 400 g
- Batteries: 3 AA batteries; typical 1000-hour life